Testing Devices

More Offers

Testing Devices • IBA Dosimetry GmbH

IBA Dosimetry · Mammo-162

Highlights

For quality assurance / acceptance test of digital Mammography Systems, according DIN 6868-162.

  • 40 mm base plate with integrated Al step wedge and 2 rows of steel balls, for checking the image limitation towards the thorax side.
  • 6 mm structural plate with recess for test inserts
  • Test insert: PMMA, SDNR & High Contrast
  • 3 × 20 mm / 1 × 10 mm / 1 × 4 mm PMMA attenuation plates
  • 1 × 20 mm PMMA full field attenuation plate (260 × 320 mm)